Results of the titration studies according to each Auto CPAP device during the whole week titration. |
|||
| Auto CPAP device |
Autoset Spirit |
Remstar Auto |
GK 420 |
|
|
|||
| CPAP usage (h) |
5.6 ± 1.9 a |
4.6 ± 1.7 a |
5.0 ± 2.3 a |
| AHI (n/h) |
7.1 ± 4.1 a |
4.7 ± 3.4 b |
3.1 ± 3.1 c |
| Max pressure (cm H2O) |
10.9 ± 1.0 a |
12.6 ± 2.4 b |
11.0 ± 2.4 a |
| Median pressure (cm H2O) |
7.4 ± 1.3 a |
6.6 ± 1.9 a |
5.9 ± 1.8 b |
| Recommended pressure level (cm H2O) |
9.8 ± 1.1 a |
9.0 ± 2.0 a |
9.2 ± 2.7 a |
| Pressure variability |
2.6 ± 0.5 a |
1.8 ± 1.0 b |
2.9 ± 1.6 a |
|
Significant differences between columns are indicated by the use of different letters as described in table 2. | |||
Sériès et al. Respiratory Research 2008 9:56 doi:10.1186/1465-9921-9-56 |
|||